Current measurement
High Current Probes: BF1, BF2, BF3-HVI-PROBE
Low Current Probes: HV-LI-PROBE
Voltage measurement
HV-U-PROBE 200V
HV-U-PROBE 1000V
Power measurement
HV-HP-PROBES, HV-LP-PROBES for simultaneous measurement of current and voltage at one input
HV-Breakout-Box
Temperature Measurement
HV-T-Probe (Thermocouple Typ K
For detailed technical information please refer to the data sheet „KLARI-PROBES“
Measurement Range of Inputs
Gain | Measuring range [mV] | Resolution [µV/Bit] |
100 | ± 9 | 0,3 |
40 | ± 27 | 0,9 |
25 | ± 42 | 1,4 |
5 | ± 210 | 7 |
1 | + 1.050/-240 | 35 |
Measurement Range HV-I-Probe
HV-I-Probe | ||||
1 mΩ | 200 μΩ | |||
Gain | Measuring range [A] | Resolution [mA/Bit] | Measuring range [A] | Resolution [mA/Bit] |
100 | ± 9 | 0,3 | ± 45 | 1,5 |
40 | ± 27 | 0,9 | ± 135 | 4,5 |
25 | ± 42 | 1,4 | ± 210 | 7 |
5 | ± 210 | 7 | ± 1.050 | 35 |
1 | + 1.050/-240 | 35 | + 5.250/-1.200 | 175 |
Measurement Range HV-U-Probe
HV-U-PROBE | ||||
200 V | 1,000 V | |||
Gain | Measuring range [V] | Resolution [mV/Bit] | Measuring range [V] | Resolution [mV/Bit] |
100 | ± 6 | 0.2 | ± 45 | 1.5 |
40 | ± 18 | 0.6 | ± 135 | 4.5 |
25 | ± 28 | 0.9 | ± 210 | 7 |
5 | ± 140 | 5 | ± 1,000 | 35 |
1 | -160/+700 | 24 | - | - |
8-channel high-voltage measurement module with 4 universal probe inputs for current, voltage, power or temperature on potentials up to 1,500 V.
Calculation of work, power, rms values, frequency, power factor, charge, discharge and total balance.
Measurement data output via Ethernet
Features
Version
Accessories
Applications
Scope of delivery
Technical data
Inputs | 4 Inputs - 8 Channel (2 x ADCs per Input) |
Capabilities | Klaric HV-Probes with automatic detection and import of the calibration values Current, Voltage, 2xVoltage, Power, HV Thermocouple Type K, HV PT100/1000 Klaric HV-Breakout Boxes |
Resolution | 16 Bit ADC with 5 measurement ranges |
Sample rate | 0,25 Hz to 8 kHz per channel configurable, dynamic sampling speed trigger |
Measurement ranges | ±9 mV, ±27 mV, ±42 mV, ±210 mV, +1050 / -240 mV 0,3 µV, 0,9 µV, 1,4 µV, 7 µV, 35 µV Resolution |
Accuracy | ± 0,1 % reading ± 3 Bit of the actual measurement range at 23°C ± 5°C ± 1 % reading ± 3 Bit of the actual measurement range -40°C bis +80°C Measurement Modules + Klari-Probes in a chain |
Ethernet | 100 MBit/s (100 BASE-TX) |
Configuration via | Ethernet |
Voltage supply range | 7 V - 60 V |
Power consumption | typ. 1 W |
Temperature range | -40°C to 85°C (-40°F to 185°F) |
Calibration | 12 months |
HV Insulation check | at least 12 months |