Temperature measurement
HV-4T-R2P-Probe (Thermocouple Type K)
For detailed technical information please refer to the data sheet „KLARI-PROBES“
Measurement range of input
Gain | Measuring range [mV] | Resolution [µV/Bit] |
100 | ± 9 | 0,3 |
40 | ± 27 | 0,9 |
25 | ± 42 | 1,4 |
5 | ± 210 | 7 |
1 | + 1.050/-240 | 35 |
4 channel HV measurement module with probe input for thermocouples.
Data output via CAN
Features
Version
Accessories
Applications
Scope of delivery
Technical data
Input | 1 input 4 channels |
Capabilities | HV-Thermocouple Type K |
Cold Junction Compensation | for each channel separate |
Resolution | 16 Bit ADC with 5 measurement ranges |
Sample rate | 0,25 Hz to 8 kHz per channel configurable, dynamic sampling speed trigger |
Measurement ranges | ±9 mV, ±27 mV, ±42 mV, ±210 mV, +1050 / -240 mV 0,3 µV, 0,9 µV, 1,4 µV, 7 µV, 35 µV Resolution Depending on temperature sensor (see specification) |
Accuracy | ± 0,1 % reading ± 3 Bit of the actual measurement range at 23°C ± 5°C ± 1 % reading ± 3 Bit of the actual measurement range -40°C bis +80°C Measurement Modules + Klari-Probes in a chain |
CAN interface | 125k, 250k, 500k, 1000k Baud configurable internal CAN termination via Software switchable CAN Base ID configurable |
Configuration via | CAN |
Voltage supply range | 7 V - 60 V |
Power consumption | typ. 1 W |
Temperature range | -40°C to 85°C (-40°F to 185°F) |
Calibration | 12 months |
HV Insulation check | at least 12 months |