Current measurement
High Current Probes: BF1, BF2, BF3-HVI-PROBE
Low Current Probes: HV-LI-PROBE
Voltage measurement
HV-U-PROBE 200V
HV-U-PROBE 1000V
Power measurement
HV BREAKOUT BOX
HV-HP-PROBES
HV-LP-PROBES
Please refer to the data sheets for detailed technical information
Measurement Range of Inputs
Gain | Measuring range [mV] | Resolution [µV/Bit] |
100 | ± 9 | 0,3 |
40 | ± 27 | 0,9 |
25 | ± 42 | 1,4 |
5 | ± 210 | 7 |
1 | + 1.050/-240 | 35 |
Measurement Range HV-I-Probe
HV-I-Probe | ||||
1 mΩ | 200 μΩ | |||
Gain | Measuring range [A] | Resolution [mA/Bit] | Measuring range [A] | Resolution [mA/Bit] |
100 | ± 9 | 0,3 | ± 45 | 1,5 |
40 | ± 27 | 0,9 | ± 135 | 4,5 |
25 | ± 42 | 1,4 | ± 210 | 7 |
5 | ± 210 | 7 | ± 1.050 | 35 |
1 | + 1.050/-240 | 35 | + 5.250/-1.200 | 175 |
Measurement Range HV-U-Probe
HV-U-PROBE | ||||
200 V | 1,000 V | |||
Gain | Measuring range [V] | Resolution [mV/Bit] | Measuring range [V] | Resolution [mV/Bit] |
100 | ± 6 | 0.2 | ± 45 | 1.5 |
40 | ± 18 | 0.6 | ± 135 | 4.5 |
25 | ± 28 | 0.9 | ± 210 | 7 |
5 | ± 140 | 5 | ± 1,000 | 35 |
1 | -160/+700 | 24 | - | - |
8-channel high-voltage measurement module with 4 universal test inputs for current, voltage, power or temperature at potentials of up to 1,500 V.
Calculation of work, power, effective values, frequency, power factor, charge, discharge and total balance.
Output via CAN
Features
Fabrication
Accesories
Applications
Scope of delivery
Technical specifications
Inputs | 4 inputs 8 channels (2 x ADC per input) |
Connectable sensors | Klaric HV-Probes with automatic recognition of the calibration values current, voltage, 2 x voltage, Combi-I/U HV thermocouple type K HV PT100/1000 Klaric HV measuring adapter |
Resolution | 16 Bit ADC with 5 measurement ranges |
Sampling rate | 0,25 Hz bis 8 kHz configurable per channel, dynamically triggerable via threshold |
Measurement range | ±9 mV, ±27 mV, ±42 mV, ±210 mV, +1050 / -240 mV 0,4 µV, 0,9 µV, 1,4 µV, 7 µV, 35 µV Resolution |
Measurement accuracy | ± 0.1 % of measured value ± 3 bits per measuring range at 23°C ± 5°C ± 1 % of measured value ± 3 bits per measuring range -40°C to +80°C Measuring module + Klari probe |
CAN interface | 125k, 250k, 500k, 1000k baud configurable internal CAN termination can be activated via software CAN base ID configurable |
Configuration via | USB (virtueller COM Port) CAN Ethernet |
Voltage supply | 7 V - 60 V |
Current consumption | typ. 2 W |
Temperature range | -40°C bis +85°C |
Calibration | Factory calibration or DAkkS calibration Recommended every 12 months |
HV-Insulation check | at least every 12 months |