Current measurement
FUSE-Probe (Micro/ Mini/ ATO/ MAXI/ MCASE/ JCASE)
HI-Probe (BF1/ BF2/ BF3)
LI-Probe
Voltage measurement
U-Probe (80V)
Power measurement
HP-Probe (Option)
LP-Probe (Option)
Temperature measurement
T-Probe (Pt100/ Pt1000)
Measurement range of the inputs
Gain | Measuring range [mV] | Resolution [µV/Bit] |
100 | ± 9 | 0,3 |
40 | ± 27 | 0,9 |
25 | ± 42 | 1,4 |
5 | ± 210 | 7 |
1 | + 1.050/-240 | 35 |
Measurement range HI-Probe
HI-Probe | ||||
1 mΩ | 200 μΩ | |||
Gain | Measuring range [A] | Resolution [mA/Bit] | Measuring range [A] | Resolution [mA/Bit] |
100 | ± 9 | 0,3 | ± 45 | 1,5 |
40 | ± 27 | 0,9 | ± 135 | 4,5 |
25 | ± 42 | 1,4 | ± 210 | 7 |
5 | ± 210 | 7 | ± 1.050 | 35 |
1 | + 1.050/-240 | 35 | + 5.250/-1.200 | 175 |
Measurement range U-Probe
The table with ID 63 not exists.Compact and robust ETHERNET measurement module for low voltage applications. Housing mechanically stackable with devices for fixing in measuring environment.
Configurable functions
OPTIONS | DESCRIPTION |
BASIC | Automatic probe recognition, kombi probes, individual setting of channels and autorange function. |
DYNAMIC SAMPLING | Dynamic sampling rate switching |
BALANCE | Calculation of the charge balance (Ah) |
Features
Version
Accesories
Applications
Scope of delivery
Technical data
Inputs | 8 inputs - 16 channel |
Connectivity | Klaric LV-Probes with automatic calibration value recognition and transfer current voltage 2 x voltage (option)power (option) PT100/1000 |
Resolution | 16 Bit ADC with 5 measuring ranges |
Sampling rate | 0.25 Hz to 8 kHz per channel configurable, dynamically triggerable via threshold |
Measuring ranges | ±9 mV, ±27 mV, ±42 mV, ±210 mV, +1050 / -240 mV 0.3 µV, 0.9 µV, 1.4 µV, 7 µV, 35 µV resolution |
Measurement accuracy | ± 1% of the measured value ± 3 bits per measuring range -40°C to +80°C Measuring module + Klari probe |
Ethernet Interface | 100 MBit/s Ethernet (100BASE-TX) IP address adjustable or assigned via DHCP |
Ethernet Protocoll | XCP-on-Ethernet Probe-Logger-Protokoll (PLP) Capture Module Protocol (CMP) |
Configuration via | Ethernet |
Voltage supply | 7V - 60V |
Current consumption | typ. 1,5 W |
Termperature range | -40°C to +85°C (120°C optional) |
Calibration | every 12 months (recommended) |