Current measurement
FUSE-Probe (Micro/ Mini/ ATO/ MAXI/ MCASE/ JCASE)
HI-Probe (BF1/ BF2/ BF3)
LI-Probe
Voltage measurement
U-Probe (80V)
Power measurement
HP-Probe (Option)
LP-Probe (Option)
Temperature measurement
T-Probe (Pt100/ Pt1000)
Measurement range of the inputs
Gain | Measuring range [mV] | Resolution [µV/Bit] |
100 | ± 9 | 0,3 |
40 | ± 27 | 0,9 |
25 | ± 42 | 1,4 |
5 | ± 210 | 7 |
1 | + 1.050/-240 | 35 |
Measurement range HI-Probe
HI-Probe | ||||
1 mΩ | 200 μΩ | |||
Gain | Measuring range [A] | Resolution [mA/Bit] | Measuring range [A] | Resolution [mA/Bit] |
100 | ± 9 | 0,3 | ± 45 | 1,5 |
40 | ± 27 | 0,9 | ± 135 | 4,5 |
25 | ± 42 | 1,4 | ± 210 | 7 |
5 | ± 210 | 7 | ± 1.050 | 35 |
1 | + 1.050/-240 | 35 | + 5.250/-1.200 | 175 |
Measurement range U-Probe
The table with ID 63 not exists.Compact and robust CAN measurement module for low voltage applications. Housing mechanically stackable with devices for fixing in measuring environment.
Features
Version
Accesories
Applications
Scope of delivery
Technical data
Inputs | 4 Measurement inputs (8 channels) |
Capabilities | Klaric Probes with automatic recognition and acceptance of calibration values Current Voltage 2xVoltage (Option) Power (Option) PT100/1000 |
Resolution | 16 bit ADC with 5 measuring ranges |
Sample Rate | 0,25 Hz bis 8 kHz configurable per channel |
Measurement Ranges | ±9 mV, ±27 mV, ±42 mV, ±210 mV, +1050 / -240 mV 0,4 µV, 0,9 µV, 1,4 µV, 7 µV, 35 µV resolution |
Accuracy | ± 0,1 % reading ± 3 Bit of the actual measurement range at 23°C ± 5°C ± 1 % reading ± 3 Bit of the actual measurement range -40°C bis +80°C |
CAN Interface | 125k, 250k, 500k, 1000k Baud configurable Internal CAN termination can be activated via software CAN base ID configurable |
Configuration via | CAN |
Voltage Supply Range | 7V - 60V |
Power Consumption | typ. 1,5 W |
Temperature Range | -40°C bis +85°C |
Calibration | Every 12 months (recommended) |