Temperature
HV-T-Probe (Thermocouple Typ K)
4-channel high-voltage measuring module with 1 inputs for temperature measurements on potentials up to 1,500 V.
Measurement and output of temperature with HV thermocouples type K.
Measurement data output via CAN.
Features
Version
Accessories
Application
Scope of delivery
Technical specifications
Inputs | 1 input (4 channels Redel 2P) 4 inputs (1 channel per input Redel 2P) |
Connectivity | Klaric HV-T-Probe with automatic detection and acceptance of the calibration values (HV thermocouple type K) |
Cold junction compensation | Separately for each channel |
Resolution | 16-bit ADC with 5 measuring ranges |
Sampling rate | 0,25 Hz bis 8 kHz pro Kanal konfigurierbar, dynamisch über Schwelle triggerbar |
Measuring ranges | -200°C to 1200°C with a resolution of 0.1°C (KLARI) |
Measurement accuracy | ± 0.1% of the measured value ± 3 bits per measuring range at 23°C ± 5°C ± 1% of the measured value ± 3 bits per measuring range -40°C to +80°C measuring module + Klari -Sample |
CAN interface | 125k, 250k, 500k, 1000k baud configurable internal CAN termination can be activated via software CAN base ID configurable |
Configuration via | USB (virtual COM Port)CANEthernet |
Power supply | 7 V - 60 V |
Power consumption | typ. 2 W |
Temperature range | -40°C bis +85°C |
Calibration | Factory calibration or DAkkS calibration Recommended every 12 months |
HV insulation check | at least every 12 months |