Current-Measurement
FUSE-Probe (Micro/ Mini/ ATO/ MAXI/ MCASE/ JCASE)
HI-Probe (BF1/ BF2/ BF3)
LI-Probe
Voltage-Measurement
U-Probe (80 V)
Power-Measurement
HP-Probe
LP-Probe
Temperature-Measurement
T-Probe (Pt100/ Pt1000)
Measuring ranges of the inputs
Gain | Measuring range [mV] | Resolution [µV/Bit] |
100 | ± 9 | 0,3 |
40 | ± 27 | 0,9 |
25 | ± 42 | 1,4 |
5 | ± 210 | 7 |
1 | + 1.050/-240 | 35 |
Measuring ranges HI-Probe
HI-Probe | ||||
1 mΩ | 200 μΩ | |||
Gain | Measuring range [A] | Resolution [mA/Bit] | Measuring range [A] | Resolution [mA/Bit] |
100 | ± 9 | 0,3 | ± 45 | 1,5 |
40 | ± 27 | 0,9 | ± 135 | 4,5 |
25 | ± 42 | 1,4 | ± 210 | 7 |
5 | ± 210 | 7 | ± 1.050 | 35 |
1 | + 1.050/-240 | 35 | + 5.250/-1.200 | 175 |
Measuring ranges U-Probes
The table with ID 63 not exists.16 channel current measurement module with 8 universal probe inputs for current, voltage, power or temperature. Calculation of charge, discharge and total balance. Output via CAN and Ethernet (XCP-on-Ethernet or KlaricServer) possible.
Features
Version
Accessories
Applications
Scope of Delivery
Technical Data
Inputs | 8 Inputs 16 Channels (2 x ADC per Input) |
Connectivity | Klaric LV probes with automatic detection and acceptance of the calibration values current voltage 2xvoltage combi-I/U PT100/1000 |
Resolution | 16-bit ADC with 5 measuring ranges |
Sampling rate | 0.25 Hz to 8 kHz per channel configurable, dynamically triggerable via threshold |
Measuring ranges | ±9 mV, ±27 mV, ±42 mV, ±210 mV, +1050 / -240 mV 0.4 µV, 0.9 µV, 1.4 µV, 7 µV, 35 µV resolution |
Measurement accuracy | ± 0.1% of the measured value ± 3 bits per measuring range at 23°C ± 5°C ± 1% of the measured value ± 3 bits per measuring range -40°C to +80°C measuring module + Klari -Sample |
CAN interface | 125k, 250k, 500k, 1000k baud configurable internal CAN termination can be activated via software CAN base ID configurable |
Configuration via | USB (virtual COM Port) CAN Ethernet |
Power supply | 7V - 60V |
Power consumption | typ. 2 W |
Temperature range | -40°C bis +85°C |
Calibration intervals | Recommended every 12 months |