Current measurement
High current-probes: BF1, BF2, BF3-HVI-PROBE
Low current-probes: HV-LI-PROBE
Voltage measurement
HV-U-PROBE 200V
HV-U-PROBE 1000V
Current/ voltage measurement
HV-I/U-COMBI-PROBE for simultaneous measurement of current and voltage at an input
Please refer to the data sheets for detailed technical information
HV-I-PROBE
HV-I-Probe | ||||
1 mΩ | 200 μΩ | |||
Gain | Measuring range [A] | Resolution [mA/Bit] | Measuring range [A] | Resolution [mA/Bit] |
100 | ± 9 | 0,3 | ± 45 | 1,5 |
40 | ± 27 | 0,9 | ± 135 | 4,5 |
25 | ± 42 | 1,4 | ± 210 | 7 |
5 | ± 210 | 7 | ± 1.050 | 35 |
1 | + 1.050/-240 | 35 | + 5.250/-1.200 | 175 |
HV-U-PROBE
HV-U-PROBE | ||||
200 V | 1000 V | |||
Gain | Measuring range [V] | Resolution [mV/Bit] | Measuring range [V] | Resolution [mV/Bit] |
100 | ± 6 | 0,2 | ± 45 | 1,5 |
40 | ± 18 | 0,6 | ± 135 | 4,5 |
25 | ± 28 | 0,9 | ± 210 | 7 |
5 | ± 140 | 5 | ± 1.000 | 35 |
1 | -160/+700 | 24 | - | - |
Measurement tasks in the vehicle, on the test bench or in the laboratory. 8 channels with 4 inputs for current, voltage or power. Measuring channels galvanically isolated. Output of performance data via CAN and Ethernet (XCP-on-Ethernet or KlaricServer) possible.
Features
Fabrication
Accesories
Applications
Scope of delivery
Technical specifications
Inputs | 4 inputs 8 channels (2 x ADC per input) |
Connectivity | Klaric HV probes with automatic detection and acceptance of the calibration values current, voltage, 2xvoltage, combination I/U, Klaric HV measuring adapter |
Resolution | 24 Bit ADC |
Sampling rate | maximum 1 MHz per channel (adjustable) |
Measuring ranges | Measuring range ±1.5 V Resolution 0.150 μV |
Measurement accuracy | ± 0.1% of the measured value ± 3 bits per measuring range at 23°C ± 5°C ± 1% of the measured value ± 3 bits per measuring range -40°C to +80°C measuring module + Klari - Probe |
Configuration via | USB (virtual COM Port) CAN Ethernet |
Voltage supply | 7 V - 60 V |
Current consumption | typ. 19 W |
Temperature range | -40°C bis +85°C |
Calibration | Factory calibration or DAkkS calibration Recommended every 12 months |
HV insulation check | at least every 12 months |